Oxygen Saturation MonitorPULSOX™-2Instruction Manual9222-1731-14_P-2.indd i9222-1731-14_P-2.indd i 09.8.20 11:03:21 AM09.8.20 11:03:21 AM
8MeasurementsMeasurements must be taken with the index fi nger placed in the fi nger holder. This instrument must be held tightly. Put the thumb only on
9Safety Notes This finger holder is designed only for use on the fingers of adults, so it should not be used on any other body parts. Never secure the
10Ending Measurements1. Open the fi nger holder and remove the fi nger.• Measurement will stop automatically, and the display will go off after 10 seco
11How to Read the Measurement ValuesPulse level meter: Indicates the pulse level in 8 levels (0 to 10%, full scale if 10% or higher). The pulse level
12Back LightThe backlight will light up automatically when the surroundings be-come dark. When backlight isn’t turned by the bright surroundings, cove
13Error Messages Messages relating to connection of the fi nger holder and attachment to the patientError MessagesThe Measure-ment Value andflash alt
14Error MessagesThe Measure-ment Value andCause Solution• Motion artifact• The pulse rate ex-ceeds the upper limit (250 bpm) of the measurement range
155. TROUBLESHOOTING Main Unit• Display disap-peared in the middle of measurements.• No display ap-pears when the finger holder was opened and the me
166. In Case of Malfunction• If a malfunction occurs, contact the nearest authorized service facility. Never disassemble the unit or attempt to rep
1710. SPECIFICATIONSPULSOX-2 Dual-wavelength pulse-type oximeter Functions:• Measuring range: SpO2: 0 to 100% Pulse rate: 20 to 250 bpm• Accura
This indicates text consisting of a warning or precaution relating to safety. Please read the text carefully and use the equipment safely.For North Am
1811. REFERENCEMeasurement PrincipleOxygen Saturation Monitor PULSOX-2 is a photometric instrument that non-invasively and continuously measures the o
19Relation between Oxygen Saturation and Partial PressureThe relation between oxygen saturation (SpO2, %) and oxygen partial pressure (PaO2, Torr or
20Guidance and manufactures declaration - electromagnetic emissionsGuidance and manufactures declaration - electromagnetic immunityThe Model PULSOX-
21Guidance and manufactures declaration - electromagnetic immunityThe Model PULSOX-2 is intended for use in the electromagnetic environment specified
9222-1731-163-91, Daisen-nishimachi, Sakai-ku, Sakai-shi, Osaka 590-8551, JapanBDEAPX Printed in Japan© 2002-2013 KONICA MINOLTA, INC.
1SAFETY PRECAUTIONSTo ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read th
21. BEFORE OPERATIONSafety Notes CAUTION: U.S. Federal law restricts this device to sale by or on the order of a physician. The PULSOX-2 is a medic
32. NAMES OF PARTS Main unitBattery cover: Open or close this cover when replacing the batteries with new ones.Battery cover lock: It is used to loc
43. NOTES ON USE• This instrument should be used in areas with an ambient temperature of 0 to 40°C (32 to 104°F) and a relative humidity of 30 to 95
5This equipment has been tested and found to comply with the limits for medical devices to the IEC 60601-1-2:2001, Medical Device Directive 93/42/EEC
64. OPERATING METHODPreparations(1) Installing Batteries(Two AAA-size batteries are required.)1. Push the battery cover, and unlock the battery cover
75. Close the battery cover lock until it clicks. • The batteries will last approxi-mately 80 hours.• The battery mark will begin to blink when th